Hoefer TE70X Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Equipamento Hoefer TE70X. Hoefer TE70X User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - Blotters

user manualum    TE70X-IM/Rev.C0/07-12TE70X Semi-dry BlottersTE70X Semi-dry transfer unitTE77X Semi-dry transfer unitTE70XP Semi-dry transfer unitTE7

Página 2 - Contents

• p2UnpackingUnwrap all packages carefully and compare contents with the packing list, making sure all items arrived. If any part is missing, contact

Página 3 - Vigtig Information – Danish

Specifications TE70X TE77X TE70XP TE77XPElectrode size (cm) 14 × 16 21 × 26 14 × 16 21 × 26Inputs: Voltage (V) 30 (DC)* 30 (DC)* 100

Página 4 - Tärkeää Tietoa – Finnish

• p42. Operating instructions To transfer proteins, prepare the unit, assemble the stack, and connect to a power supply if necessary. Then run the tra

Página 5

• p54Prepare the blotting paperFor each gel, cut at least 6 pieces of blotting paper the same size as the gel or slightly smaller. Gauge the thickness

Página 6 - Portuguese

• p66Complete the stacka. Place the pre-wet membrane onto the stack of blotting paper.b. Place the gel on the membrane. Note: Proteins bind to the

Página 7 - Viktig Information – Swedish

• p7Multiple gels: Either lay gels of the same thickness side-by-side (Fig 3), or stack 2 sandwiches layered as shown (Fig 4).For best results, the tr

Página 8 - Electronic Equipment (WEEE)

• p8Blotting paper, (2–3 sheets)Blotting paper, (1 sheet)CellophaneMask (optional)GelMembraneBlotting paper, (2– 3 sheets)GelMembraneFig 4. Transfer s

Página 9 - Description

• p9ElectrotransferTE70X and  TE77X1The power supply should be switched off and both the current and voltage controls set at zeroThen plug the color-c

Página 10 - Unpacking

• p10TE70XP and  TE77XP1Turn the instrument on using the POWER button on the front keypad. The instrument works by setting the current and time, and t

Página 11 - Specifications

• p11power ON/OFFDuring the  transferTE70XP  and the TE77XPA red LED will appear in the lower right corner of the Display when voltage is applied to

Página 12 - 2. Operating instructions 

• piContents Important Information ... iiWaste Electrical and Electronic Equipment (WEEE) ...

Página 13

• p12After the  transfer is completeTE70X  and TE77XTurn off the power supply and disconnect the leads from the power supply.TE70XP  and TE77XPAfter a

Página 14

• p133. Care and maintenance• Do not autoclave or wash the unit in a dishwasher.• Do not immerse the unit in water!Rinse the cathode in the cover an

Página 15 - Membrane

• p144. Troubleshootingproblem  solution Incomplete transfer Blank or faint areas on Remove trapped air pockets between the gel andthe membrane

Página 16 - Fig 4. Transfer stack for

• p15problem  solution Open circuit or no output current  Push circuit protection reset button located next tofor TE70X or TE77X   the input leads. To

Página 17 - Electrotransfer

• p16problem  solution Inefficient binding to membrane Chemical parameters Prepare transfer buffer without SDS. (SDS can improve transfer efficiency

Página 18 - TE70XP and  TE77XP

• p175. Electrotransfer notes• Run the transfer as soon as possible after electrophoresis to minimize protein diffusion within the gel.• Stacked g

Página 19 - During the  transfer

• p18CAPS  buffer, 1X(10 mM CAPS, pH 11.0, 1 liter)CAPS (FW 221.3) 10 mM 2.2 g [3-(cyclohexylamino)-1-propanesulfonic acid]Dissolve in 600 ml dist

Página 20

• p196. BibliographyBjerrum, O.J., Larsen, K., and Heegaard, N., CRC Handbook of Immunoblotting of Proteins 1, Section 7. CRC Press (1988).Gallagher,

Página 21 - 3. Care and maintenance

• p207. Customer service informationTechnical service and repairHoefer, Inc. offers complete technical support for all of our products. If you have an

Página 22 - 4. Troubleshooting

• p218. Ordering informationproduct   qty.  code no.TE70X Semi-dry Transfer Unit, 14 × 16 cm  1 TE70X Includes 25 sheets of blotting paper, 50 sh

Página 23

• piiImportant Information  – English• If this equipment is used in a manner not specied by Hoefer, Inc. the protection provided by the equipment ma

Página 24

• p22Related productsproduct     code no.SE600 Chroma SE600X-15-1.5MiniVE Vertical Electrophoresis system SE300-10A-1.0PBS300B Powe

Página 25 - 5. Electrotransfer notes

• p23Hoefer, Inc. 84 October Hill Road Holliston, MA 01746Toll Free: 1-800-227-4750 Phone: 1-508-893-8999 Fax: 1-508-893-0176 E-mail: support@hoeferin

Página 26

• piii80%). Undgå kondensation ved lade enheden equili-brate til omgivende temperatur ved tageen instru-mentets fra et koldere til et varmere miljø. •

Página 27 - 6. Bibliography

• pivil devrait être seulement utilisé par les opérateurs convenablement entraînés. Lire ce manuel entier avant utiliser l’instrument et l’usage seule

Página 28 - Technical service and repair

• pvsopra di 80%). Evitare la condensazione lasciando l’unità equilibra alla temperatura ambiente quando portare lo strumento da un più freddo a un am

Página 29 - 8. Ordering information

• pvi• Este instrumento é projectado de acordo com o EN61010-1:2001 condição de segurança eléctrica. Não obstante, deve ser usado só por operadores a

Página 30 - Related products

• pviiEnglishFrenchGermanItalianSpanishSwedishThis symbol indicates that the waste of electrical and electronic equipment must not be disposed as unso

Página 31

1. TE70X Semi-dry Blotters:  DescriptionThe TE70X, TE77X, TE70XP, and the TE77XP semi-dry blotters rapidly transfer proteins from polyacrylamide gels